Appendix E: QC Using BD FACSDiva Software 325
e Reset the window extension to the appropriate setting (typically 2).
3 Adjust the red laser area scaling until the APC-A signal matches the APC-H
signal, if needed.
Click the Laser tab in the Cytometer window.
•Increase the area scaling factor if the APC-A signal is lower than
APC-H.
•Decrease the area scaling factor if the APC-A signal is higher than
APC-H.
4 Repeat step 1 through step 3 to adjust area scaling and laser delay for the
violet (or Near UV) laser, using the appropriate parameter instead of APC.
Recording QC Data
The QC experiment template includes Browser elements, as well as plots and
instrument settings. The first time you use the QC experiment template, adjust
the settings by placing the signals for each parameter to channel 100,000. The
instrument settings you establish will be used for subsequent QC runs.
1 Display the QC worksheet by clicking the QC tab in the Worksheet
window.