Chapter 1: Cytometer Components 21
Power and Operation
Power to the fluidics cart is supplied by the cytometer. The cart is activated when
the cytometer main power switch is turned on (see Power Panel on page 39).
Figure 1-5 Fluidics cart flow gauge
See the following sections for more information about the fluidics cart:
• Setting Up the Fluidics Cart on page 122
•Refilling the Plastic Containers on page126
• Emptying the Waste Container on page 127
Power to the fluidics cart is supplied and controlled through the flow
cytometer. The fluidics cart voltage settings have been configured to match
the supply voltage by your service engineer. To properly operate the fluidics
cart, plug the fluidics cart power cord only into the power receptacle on the
cytometer (Figure 1-3 on page 20). Do not plug the power cord directly
into a wall socket. Do not change the input voltage selection switch on the
fluidics cart.
When the stream is on, air pressure fluctuates between 80–100 psi
(Figure 1-5). A pressure reading of less than 80 psi or greater than 100 psi
indicates that the fluidics cart is not functioning properly. If this occurs,
contact your BD Biosciences service representative for assistance. Do not
operate the cytometer outside the normal air pressure range.
pressure gauge
at approximately
88 psi