134 BD FACSAria II User’s Guide
Running a Performance Check
The performance check feature of Cytometer Setup and Tracking will check the
cytometer’s daily performance.
1 Install the bead tube onto the cytometer loading port.
2 In the Setup Control window, select Check Performance from the
Characterize menu.
3 Click Run.
Plots appear under the Setup tab and the performance check is run. The
performance check takes approximately 5 minutes to complete.
Reviewing the Results
1 Once the performance check is complete, click View Report.
2 Print the Report.
3 Verify that the performance check passed.
In the Setup tab, the Cytometer Performance Results should have a green
checkmark and the word Passed next to it.
If any parameters did not pass, refer to The Cytometer Setup and Tracking
Application Guide for troubleshooting information.
Komentáře k této Příručce